2.2. ELECTRICITY SUPPLY
The electrical installation must comply with the current standards in the country.
The electricity supply must be single phase 230 volts ±10%, 50/60Hz, 10A.
Installation category / Mains overload category = II
An earth connection is essential.
The maximum absorbed power of the sterilizer is 2100 W (9,2A).
The installation must include:
- an earthed plug.
- a 10A differential circuit breaker with a sensitivity of 30mA (2).
Plug in the mains cable to the back of the sterilizer.
Check that the serviceable voltage specified on the name plate located on the back of the
sterilizer
corresponds to the supplied mains voltage.
The overall electrical safety of the sterilizer is only guaranteed if the
sterilizer
is appropriately earthed.
If unclear, have the installation checked by a qualified electrician.
Do not plug other equipment into the same socket.
Do not bend or twist the mains cable.
The sterilizer must be connected to an electrically earthed plug.
Use only the mains cable delivered with the sterilizer.
2
+
2
+
W&H l
STERILIZATION
does not accept any responsibility if these instructions are not complied with.