2. SET-UP
2.1. INSTALLATION
The sterilizer has been calibrated and tested in the factory. It does not require calibration during installation.
Working temperature range: 10°C to 40°C / Humidity : 0 to 90%.
Storage temperature range : -20°C to 60°C / Humidity : 0 to 90% (Water tanks empty).
Install the sterilizer as outlined below:
Install the sterilizer on a flat and level surface.
The maximum weight of the sterilizer with the main water tank full and the chamber fully loaded is:
57,0 kg, 129.4 N/m², 140 N/foot (MB17)
66.5 kg, 177.3 N/m², 160 N/foot (MB22)
Leave a gap of 5 cm at the back and 2 cm on each side of the sterilizer to ensure adequate ventilation.
Do not place the sterilizer near a sink or in a location where it is likely to be splashed.
Install the sterilizer in a well ventilated room.
Keep away from all sources of heat.