7.6. CLEANING THE WATER TANKS
Completely drain both water tanks (main and used water tank).
Open the service door.
Insert the drain tube into the quick coupling drain connection of
the used water tank (right). Allow the entire contents of the tank
to empty and discard the used water.
Disconnect the drainage hose by pressing the push-button on the
drain connection.
Insert the drain tube into the quick coupling drain connection of
the main water tank (left). Allow the entire contents of the tank to
empty.
Disconnect the drainage hose by pressing the push-button on the
drain connection.
Fill both reservoirs each with:
2.8 litres of demineralised water + 0.2 litres of 90% alcohol (MB17)
3.7 litres of demineralised water + 0.3 litres of 90% alcohol (MB22)
Allow the solution to sit for 30 minutes.
Drain both tanks and discard the 3 litres on MB17 (4 litres on MB22) of solution.
Fill the main tank with 3 litres on MB17 (4 litres on MB22) of distilled or
demineralised water.
Run an empty cycle.
If the sterilizer is not used for more than 3 days, both water tanks must be
completely drained in order to avoid alga growth or any other deposits.
DO NOT RUN A CYCLE !
+