User manual
31
User manual
7.2. CLEANING THE DOOR SEAL
Clean the door seal and the porthole with a lint free cloth saturated with alcohol.
The porthole can also be cleaned with a non-abrasive detergent.
7.3. CLEANING THE CHAMBER, TRAYS AND TRAY HOLDER
Remove the trays from the chamber.
Disconnect and remove the rack.
Clean the chamber with a damp sponge moistened with a detergent or scouring agent if necessary.
Rinse with a damp sponge to remove all traces of the cleaning agent.
Apply the same procedure for the rack, trays (cassettes).
7.4. CLEANING THE EXTERNAL PARTS
Clean the external parts with a damp cloth and mild detergent.
Never use scouring agents or highly abrasive products.
7.5. REPLACEMENT OF THE BACTERIOLOGICAL FILTER
Open the service door.
Unscrew the bacteriological filter by hand (anti-clockwise).
Insert and manually screw the new filter into position
Ensure that you clean all around the sterilizer chamber.
Do not bend or damage the temperature sensor at the bottom of the chamber.
Never use disinfectants to clean the chamber.
Do not use copious amounts of water to wash the sterilizer as this may damage the
electrical components and safety mechanisms.
Take care not to scratch the plastic film in front of the touch-screen.