7. USER MAINTENANCE
A distinction must be made between two levels of maintenance, that performed regularly by the user and preventive
maintenance carried out by an approved technician (§ 8).
7.1. MAINTENANCE PROGRAM
Frequency /
N° of
cycles
Operation
Spare
number
Description
Weekly
50
Cleaning the door seal.
Cleaning the chamber, the trays and the rack.
Cleaning the external surfaces.
-
-
-
§ 7.2.
§ 7.3.
§ 7.4.
Every 3 months
400 Replacing the bacteriological filter.
W322400X § 7.5.
Every 6 months
1000 Cleaning both water tanks.
- § 7.6.
Every year
1000 Replacing the door seal.
F460503X § 7.7.
* Every 3 years.
4000 Servicing by an approved technician.
- § 8.
*Refer to the legislation and instructions for each individual country.
The maintenance sub-menu 4.5.7.) indicates the number of cycles remaining before replacement of the
bacteriological filter, the door seal and when general servicing is required. The three counters are decreased in value
after each cycle.
When one of the counters reaches 0 a corresponding message appears at the bottom of the selection screen.
It is not possible to run a new cycle (the selection button will disappear) unless the reading of the message has been
confirmed by pushing the "OK" icon. The counter is then automatically reset.
If one of the three operations is completed before the respective counter has reached 0 it is necessary to reset the
counter manually. Place the cursor in front of the operation with the icons UP and DOWN and reset it by pushing the
"OK" icon.
Remove the mains cable before examining the sterilizer.
Cursor UP
Cursor DOWN
Reset of the
selected
counter
Exit of the sub-menu