User manual

29
User manual
DESCRIPTION OF THE ALARMS (continue)
Description Action
Vacuum pump
A31
During a vacuum phase, the maximum achieved
vacuum is not lower than – 0.20 bar.
Clean and check the door seal (§ 7.2.), repeat the
cycle. If the problem persists
service.
A32
During a vacuum phase, the maximum achieved
vacuum is not lower than – 0.50 bar.
Clean and check the door seal (§ 7.2.), repeat the
cycle. If the problem persists
service.
A33
The global vacuum level obtained after 5 pulses is
not sufficient. The calculated additional 6th pulse is
impossible to obtain (outside limits).
Clean and check the door seal (§ 7.2.), repeat the
cycle. If the problem persists
service.
A34
The last 10 cycles have required a 6th additional
vacuum pulse.
Sterilization is guaranteed because the additional
6th pulse ensures the global vacuum level has been
obtained,
service for control.
Door locking
A52
1) The door locking system is blocked during
opening and closing phases.
2) The door locking switch is open during the
cycle.
service
Electro-valves
A63
The pressure does not exceed – 0.70 bar,
2 minutes after the end of a vacuum phase.
service