6. ALARMS
The micro-processor continuously analyses all the cycle parameters.
If there is any doubt regarding sterilization efficiency, the cycle is immediately interrupted and an alarm displayed.
This is followed by a phase of approximately 2 minutes required to reinitialise the sterilization system and to return the
chamber to atmospheric pressure. The load is not sterile and the cycle must be repeated !
The alarm screen appears… Press the "OK" icon at the end of the re-initialisation
phase to confirm the reading of the alarm.
Opening of the door… …and return to stand-by mode.
L O A D N O T
P L E A S E W A I T
S T E R I L E
Code of the alarm /
interrupted phase
see table §6.1.
If the cycle is interrupted after the PR phase (plateau period), following message will be
displayed and printed : !! LOAD STERILIZED, BUT NOT DRIED !!
!! FOR IMMEDIATE USE ONLY !!