5. MESSAGES
A certain number of messages can appear at the beginning or at the end of the cycle.
This is merely information and not an alarm. The user can continue to operate the sterilizer with complete confidence.
During a cycle, the micro-processor continuously analyses all the parameters. The cycle would be interrupted immediately
and an alarm displayed, if there was a concern with the sterilization quality (see § 6).
Message display
Messages list:
Message Description Action
FILL THE
WATER TANK
The minimum water level has been reached in the
main tank.
Fill the main tank using distilled or demineralised
water as described in § 4.3.
DRAIN THE
USED WATER TANK
The maximum level of the used water tank has been
reached.
Drain the tank described in § 4.4.
CLOSE
THE DOOR
The door is not closed properly. Close the door.
DOOR LOCKING
PROBLEM
The door could not be locked properly
Control (clean) the door seal.
Monitor. ☎ Service if the message persists.
Ttheo/Tsen DIFF
IN STERL. PHASE
During sterilization phase: 2°C difference between
measured and theoretical temperatures (deducted
from the pressure).
The sterilization grade is satisfactory because, at
the beginning of the cycle, the micro-processor
evaluates the vacuum level and stops the cycle if it
is inadequate.
Monitor. ☎ Service if the message persists.
Temp FLUCTUATION
IN STERIL. PHASE
During sterilization phase: fluctuation by
+
/
-
1,5°C of
the measured and/or the theoretical temperatures
(deducted from the pressure).
The sterilization grade is satisfactory because, at
the beginning of the cycle, the micro-processor
evaluates the vacuum level and stops the cycle if it
is inadequate.
Monitor. ☎ Service if the message persists.
F I L L T H E
W A T E R T A N K