4.6.3. CYCLE PROCEDURE
The following information is displayed during the cycle:
4.6.4. COPIES OF CYCLE PRINTOUTS
It is possible to get copies of the last cycle printout prior to door opening.
Access to technical
data ➫ § 4.8
Manual stop ➫ § 4.7
Date
Time
Selected cycle
N° of the cycle
Temperature
Pressure
Current phase
Total duration
Reference to
cycle data ➫ § 4.9
END OF THE CYCLE
Opening of
the door
Date
Time
Selected cycle
N° of the cycle
Temperature
Pressure
End of the cycle
Total duration
"Load sterile"
Comment: If the door remains locked at the end of the cycle, the vacuum pump remains in operation. Every
10 minutes the pump will run for 40 seconds in order to ventilate the chamber and load preventing
the build up of moisture. This operation continues until the temperature of the chamber falls below
60°C.
Before the door is opened, it is possible to access the cycle data by simply pressing the "Info" icon
(see § 4.9.).
You can access the technical parameters at
any time by pressing the "Info" icon (see §
4.8.).
If needed, the cycle can be interrupted by
pressing the "STOP" icon for 2 seconds.
An audible tone will confirm your request.
This is followed by a 2-minute interval
necessary to reinitialise the system and
revert to atmospheric pressure in the
chamber (see § 4.7.).
At the end of the cycle, the micro-processor
analyses the temperature, pressure and
duration of all the phases in order to confirm
whether or not sterilization has occurred.
After reading the comment, you can open the
door by pressing the corresponding key.
….Select the quantity needed: 1 to 9,
Print by pressing
Press
…