4.2. DESCRIPTION OF THE WATER TANK
The sterilizer is equipped with 2 independent water tanks of equal volume - 3.5 litres (MB17) or 4,5 litres (MB22).
The left tank also called the "main tank" contains the distilled or demineralized water required for each cycle.
It is fitted with 2 level sensors, minimum (0.5 l) and maximum (3.5 l on MB17 or 4.5 l on MB22).
The right tank called the "used water tank" contains the used water collected at the end of each cycles.
It is fitted with a maximum level sensor (3.5 l on MB17 or 4.5 l on MB22).
The 2 tanks are connected to drain connections located behind the service door.
The water consumption per cycle varies depending on the type and mass of the sterilized load.
The following message is displayed on the screen when the minimum water level is reached in the main tank:
While the message is displayed, it is impossible to run another cycle.
The selection button has disappeared !
Main water tank
Main water tank
drain connection
Used water tank
drain connection
Used water
tank
The sterilizer will run for at least 8 cycles before the main tank needs refilling.
F I L L T H E
W A T E R T A N K