Specifications

TRIGGER:SOURCE TIMER (*RST default)
specify data format
FORMAT ASC,7 (*RST default)
select FIFO mode
SENSE:DATA:FIFO:MODE BLOCK may read FIFO while running
Define algorithm
ALG:DEFINE ‘ALG1’,’static float a,b,c, div, mult, sub;
if ( First_loop )
{
a=1;b=2;c=3;
writecvt( a, 10 ); writefifo( b, 11 ); writefifo( c, 12 );
}
writecvt( a / div, 13 );
writecvt( b * mult, 14 );
writecvt( c - sub, 15 );’
Pre-set coefficients
ALG:SCAL ‘ALG1’,’div’,5
ALG:SCAL ‘ALG1’,’mult’,5
ALG:SCAL ‘ALG1’,’sub’,0
ALG:UPDATE
initiate trigger system (start algorithm)
INITIATE
retrieve data
SENSE:DATA:CVT? (@10:15)
Programming the VT1419A Multifunction
Plus
Example Command Sequence
Chapter 3 87