Specifications

OUTP:CURR:AMPL 488e-6,(@132:139)
set excite current to 488 µA on current SCP channels 32 through 39
SENS:FUNC:TEMP THER, 2250, (@100:107)
link channels 0 through 7 to temperature EU conversion for 2,250W thermistor
To set channels 8 through 15 to measure temperature using 10,000 W thermistors
(in this case paired to current source SCP channels 40 through 47):
OUTP:CURR:AMPL 30e-6,(@140:147)
set excite current to 30 µA on current SCP channels 40 through 47
SENS:FUNC:TEMP THER, 10000, (@108:115)
link channels 8 through 15 to temperature EU conversion for 10,000 W
thermistor
To set channel 7 to measure temperature using 100 W RTD with a TC of 0.00385
ohm/ohm/°C (in this case paired to current source SCP channel 39):
OUTP:CURR:AMPL 488e-6,(@139)
set excite current to 488 µA on current SCP channels 32 through 47
SENS:FUNC:TEMP RTD, 85, (@107)
link channel 7 to temperature EU conversion for 100 W RTDs with 0.00385
TC.
Thermocouple Measurements
Thermocouple measurements are voltage measurements that the EU conversion
changes into temperature values based on the <sub_type> parameter and latest
reference temperature value.