Specifications

NOTE At Power-on and after *RST, the default EU Conversion is autorange voltage for
analog input channels.
Linking Voltage
Measurements
To link channels to the voltage conversion send the [SENSe:]FUNCtion:VOLTage
[<range>,] (@<ch_list>) command.
·
The <ch_list> parameter specifies which channels to link to the voltage EU
conversion.
·
The optional <range> parameter can be used to choose a fixed A/D range.
Valid values are: 0.0625, 0.25, 1, 4, 16, or AUTO. When not specified, the
module uses auto-range (AUTO).
To set channels 0 through 15 to measure voltage using auto-range:
SENS:FUNC:VOLT AUTO,(@100:115)
To set channels 0 and 23 to the 16 volt range and 28 through 31 to the 0.0625 volt
range:
SENS:FUNC:VOLT 16,(@100,123)
SENS:FUNC:VOLT .625,(@128:131) must send a command per range
Note When using manual range in combination with amplifier SCPs, the EU conversion
will try to return readings which reflect the value of the input signal. However, the
user must choose range values that will provide good measurement performance
(avoiding over-ranges and select ranges that provide good resolution based on the
input signal). In general, measurements can be made at full speed using auto-range.