Specifications

Possible Corrective Action by Failed Test ID Number
Test ID Corrective Actions
20, 30 -37 Remove all SCPs and see if *TST? passes. If so,
replace SCPs one at a time until the one causing
the problem is found.
38 - 71 (VXI Technology Service)*
72,74 - 76, 80 - 93,
301 - 354
Re-seat the SCP that the channel number(s)
points to or move the SCP and see if the failure(s)
follow the SCP. If the problems move with the
SCP, replace the SCP.
73, 77 - 79, 94 - 99 (VXI Technology Service)*
*Must send module to a VXI Technology Service Center for
repair. Record information found in FIFO to assist the
VXI Technology Service Center in repairing the problem.
Refer to the Command Reference under *TST? for a list of
module functions tested.
NOTE During the first five minutes after power is applied, *TST? may fail. Allow the
module to warm-up before executing *TST?
3053 ‘Corrupt on board Flash memory.
3056 ‘Custom EU not loaded.’ May have erased custom
EU conversion table with *RST. May have linked channel with
standard EU after loading custom EU, this erases the custom
EU for this channel. Reload custom EU table using
DIAG:CUST:LIN or DIAG:CUST:PIEC.
3057 ‘Invalid ARM or TRIG source when S/H SCP’s enabled.’
Don’t set TRIG:SOUR or ARM:SOUR to SCP with VT1510A
or VT1511A installed.
3058 ‘Hardware does not have D32, S/H or new trigger capabilities.
Module’s serial number is earlier than 3313A00530.
3067 ‘Multiple attempts to erase flash memory failed.’
3068 ‘Multiple attempts to program flash memory failed.’
3069 ‘Programming voltage jumper not set properly.’ See Disabling
Flash Memory Access in Chapter 1 (JM2201).
3070 ‘Identification of Flash ROM incorrect.
3071 ‘Checksum error on flash memory.’
3074 ‘WARNING! Old Opt 16 or Opt 17 card can damage SCP
modules’ must use VT1506A or VT1507A.
Error Messages
362 Appendix B
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com