Specifications
Planning for
Thermocouple
Measurements
Thermocouples and the thermocouple reference temperature sensor can be wired to
any of the VT1419A’s channels. When the scan list is executed, make sure that the
reference temperature sensor is specified in the channel sequence before any of the
associated thermocouple channels (see the [SENSe:]REF:CHAN command).
External wiring and connections to the VT1419A are made using the Terminal
Module (see page 39).
NOTE The isothermal reference temperature measurement made by a VT1419A applies
only to thermocouple measurements made by that instrument and through the
terminal blocks associated with the reference temperature sensor (for increased
isothermal reference accuracy the VT1586A Rack Mount Terminal Panel has three
reference temperature thermistors). In systems with multiple VT1419As, each
instrument must make its own reference measurements. The reference measurement
made by one VT1419A can not be used to compensate thermocouple measurements
made by another VT1419A.
Field Wiring
Planning the Wiring Layout
28 Chapter 2
Or a Single
VT1518A
For 4 Channels
sense Hi
VT1505A
SCP
(source)
Note
Each channel line
represents both a
Hi and Lo signal
Sense
SCP
Ch 24
Ch 31
Ch 32
Ch 39
Faceplate Conns
or
T
e
rmin
a
lM
odu
l
e
sense Lo
Figure 2-2: Pairing Source and Sense SCP Channels