Specifications

·
The <channel> parameter must specify a single channel only.
·
Related Commands: FUNC:STRAIN¼, STRAIN:POISSON
Usage STRAIN:POISSON? (@131) query for the Poisson ratio specified for sense
channel 31
enter statement here enter the Poisson ratio value
[SENSe:]STRain:UNSTrained
[SENSe:]STRain:UNSTrained <unstrained_v>,(@<ch_list>) specifies the
unstrained voltage value to be used to convert strain bridge readings for the
channels specified by <ch_list>. This command does not control the output voltage
of any source.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
unstrained_v numeric (flt32) -16 through +16 volts
ch_list channel list (string) 100 - 163 none
Comments
·
Use a voltage measurement of the unstrained bridge sense channel to determine
the correct value for <unstrained_v>.
·
The <ch_list> parameter must specify the channel used to sense the bridge
voltage, not the channel position on a Bridge Completion SCP.
·
Related Commands: SENSE:STRAIN:UNST?, SENSE:FUNC:STRAIN…
·
*RST Condition: Unstrained voltage is zero.
Usage STRAIN:UNST .024,(@100) set unstrained voltage for channel 0
[SENSe:]STRain:UNSTrained?
[SENSe:]STRain:UNSTrained? (@<channel>) returns the unstrained voltage
value currently set for the sense channel specified by <channel>. This command
does not make a measurement.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel channel list (string) 100 - 163 none
Comments
·
Returned Value: Numeric value of unstrained voltage. The C-SCPI type is flt32.
·
The <channel> parameter must specify a single channel only.