Specifications
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel channel list (string) 100 - 163 none
Comments
·
Returned Value: Numeric value of gage factor. The C-SCPI type is flt32.
·
The <channel> parameter must specify a single channel only.
·
Related Commands: STRAIN:GFACTOR
Usage STRAIN:GFAC? (@107) query gage factor for channel 7
enter statement here returns the gage factor set by STR:GFAC
[SENSe:]STRain:POISson
[SENSe:]STRain:POISson <poisson_ratio>,(@<ch_list>) sets the Poisson ratio to
be used for EU conversion of values measured on sense channels specified by
<ch_list>.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
poisson_ratio numeric (flt32) 0.1 - 0.5 none
ch_list channel list (string) 100 - 163 none
Comments
·
The <ch_list> parameter must specify channels used to sense strain bridge output,
not channel positions on a Bridge Completion SCP.
·
Related Commands: FUNC:STRAIN…, STRAIN:POISson?
·
*RST Condition: Poisson ratio is 0.3
Usage STRAIN:POISSON .5,(@124:131) set Poisson ratio for sense channels 24
through 31
[SENSe:]STRain:POISson?
[SENSe:]STRain:POISson? (@<channel>) returns the Poisson ratio currently set
for the sense channel specified by <channel>.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel channel list (string) 100 - 163 none
Comments
·
Returned Value: numeric value of the Poisson ratio. C-SCPI type is flt32.
VT1419A Command Reference
[SENSe]
Chapter 6 281
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com