Specifications

[SENSe:]REFerence:TEMPerature
[SENSe:]REFerence:TEMPerature <degrees_c> stores a fixed reference junction
temperature in the Reference Temperature Register. Use when the thermocouple
reference junction is kept at a controlled temperature.
NOTE This reference temperature is used to compensate all subsequent thermocouple
measurements until the register is overwritten by another SENSE:REF:TEMP value
or by scanning a channel linked with the SENSE:REFERENCE command. If used,
SENS:REF:TEMP must be executed before scanning any thermocouple channels.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
degrees_c numeric (float32) -126 to +126 none
Comments
·
This command is used to specify to the VT1419A the temperature of a controlled
temperature thermocouple reference junction.
·
When Accepted: Not while INITiated
·
Related Commands: FUNC:TEMP TC…
·
*RST Condition: Reference temperature is 0 °C
Usage SENSE:REF:TEMP 40 subsequent thermocouple conversion will
assume compensation junction at 40 °C
[SENSe:]STRain:EXCitation
[SENSe:]STRain:EXCitation <excite_v>,(@<ch_list>) specifies the excitation
voltage value to be used to convert strain bridge readings for the channels specified
by <ch_list>. This command does not control the output voltage of any source.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
excite_v numeric (flt32) 0.01 - 99 volts
ch_list channel list (string) 100 - 163 none
Comments
·
The <ch_list> parameter must specify the channel used to sense the bridge
voltage, not the channel position on a Bridge Completion SCP.
·
Related Commands: SENSE:STRAIN:…, SENSE:FUNC:STRAIN…
·
*RST Condition: 3.9 V
VT1419A Command Reference
[SENSe]
Chapter 6 279