Specifications

·
When Accepted: Not while INITiated
·
Related Commands: *CAL?, [SENSE:]STRAIN¼
·
*RST Condition: SENSE:FUNC:VOLT 0,(@100:163)
Usage FUNC:STRAIN 1,(@100:,105,107) quarter bridge sensed at channels 0, 5 and 7
[SENSe:]FUNCtion:TEMPerature
[SENSe:]FUNCtion:TEMPerature <type>,<sub_type>,[<range>,](@<ch_list>)
links channels to an EU conversion for temperature based on the sensor specified in
<type> and <sub_type>. Not
for sensing thermocouple reference temperature
(for that, use the SENS:REF <type>,<sub_type>,(@<channel>) command).
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
type discrete (string) RTD | THERmistor | TCouple none
sub_type numeric (float32)
numeric (float32)
discrete (string)
for RTD use 85 | 92
for THER use 2250 | 5000 | 10000
for TC use CUSTom|E|EEXT |
J|K|N|R|S|T
none
ohms
none
range numeric (float32) see comments V dc
ch_list channel list (string) 100 - 163 none
Comments
·
Resistance temperature measurements (RTDs and THERmistors) require the use
of Current Source Signal Conditioning Plug-Ons. The following table shows the
Current Source setting that must be used for the following RTDs and Thermistors:
Required Current
Amplitude
Temperature Sensor Types and
Subtypes
MAX (488 µA)
MIN (30 µA)
for RTD and THER,2250
for THER,5000 and THER,10000
·
The <range> parameter: The VT1419A has five ranges: 0.0625 V dc, 0.25 V dc,
1 V dc, 4 V dc, and 16 V dc. To select a range, simply specify the range value
(for example, 4 selects the4Vdcrange). If a value is specified larger than one of
the first four ranges, the VT1419A selects the next higher range (for example, 4.1
selects the 16 V dc range). Specifying a value larger than 16 generates an error.
Specifying 0 selects the lowest range (0.0625 V dc). Specifying AUTO selects
auto range. The default range (no range parameter specified) is auto range.
·
If using amplifier SCPs, set them first and keep their settings in mind when
specifying a range setting. For instance, if expected signal voltage is to be
approximately 0.1 V dc and the amplifier SCP for that channel has a gain of 8,
<range> must be set no lower than1Vdcoraninput out-of-range condition will
exist.