[SENSe:]FUNCtion:CUSTom
[SENSe:]FUNCtion:CUSTom [<range>,](@<ch_list>) links channels with the
custom Engineering Unit Conversion table loaded with the DIAG:CUST:LINEAR
or DIAG:CUST:PIECE commands. Contact a VXI Technology System Engineer for
more information on Custom Engineering Unit Conversion for specific applications.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
range numeric (float32) see first comment V dc
ch_list channel list (string) 100 - 163 none
Comments
·
The <range> parameter: The VT1419A has five ranges: 0.0625 V dc, 0.25 V dc, 1 V dc,
4 V dc, and 16 V dc. To select a range, simply specify the range value (for example, 4
selects the 4 V dc range). If a value is specified larger than one of the first four ranges,
the VT1419A selects the next higher range (for example, 4.1 selects the 16 V dc range).
Specifying a value larger than 16 causes an error -222 “Data out of range.” Specifying 0
selects the lowest range (0.0625 V dc). Specifying AUTO selects auto range. The
default range (no range parameter specified) is auto range.
·
If using amplifier SCPs, set them first and keep their settings in mind when specifying a
range setting. For instance, if the expected signal voltage is to be approximately 0.1 V dc
and the amplifier SCP for that channel has a gain of 8, <range> must be set no lower
than 1 V dc or an input out-of-range condition will exist.
·
If an A/D reading is greater than the <table_range> specified with
DIAG:CUSTOM:PIEC, an overrange condition will occur.
·
If no custom table has been loaded for the channels specified with SENS:FUNC:CUST,
an error will be generated when an INIT command is given.
·
When Accepted: Not while INITiated
·
Related Commands: DIAG:CUST:…
·
*RST Condition: all custom EU tables erased
Usage program must put table constants into array table_block
DIAG:CUST:LIN 1,table_block,(@116:123) send table to VT1419A for chs 16-23
SENS:FUNC:CUST 1,(@116:123) link custom EU with chs 16-23
INITiate then TRIGger module