Specifications

Comments
·
When Accepted: Not while INITiated
·
Related Commands: SENSE:DATA:FIFO¼
·
*RST Condition: SENSE:DATA:FIFO:RESET
Usage SENSE:DATA:FIFO:RESET Clear the FIFO
[SENSe:]FREQuency:APERture
[SENSe:]FREQuency:APERture <gate_time>,<ch_list> sets the gate time for
frequency measurement. The gate time is the time period that the SCP will allow for
counting signal transitions in order to calculate frequency.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
gate_time numeric (float32) 0.001 to 1 (0.001 resolution) seconds
ch_list string 132 - 163 none
Comments
·
If the channels specified are on an SCP that doesn’t support this function, an error
will be generated. See the SCP’s User’s Manual for its capabilities.
·
Related Commands: SENSe:FUNCtion:FREQuency
·
*RST Condition: 0.001 seconds
Usage SENS:FREQ:APER .01,(@148) set channel 48 aperture to 10 ms
VT1419A Command Reference
[SENSe]
266 Chapter 6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com