Specifications
Usage DATA:FIFO:MODE? Enter statement returns either BLOCK or
OVERWRITE
[SENSe:]DATA:FIFO:PART?
[SENSe:]DATA:FIFO:PART? <n_values> returns n_values from the FIFO
buffer.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
n_values numeric (int32) 1 - 2,147,483,647 none
Comments
·
Use the DATA:FIFO:COUNT? command to determine the number of values in
the FIFO buffer.
·
The format of values returned is set using the FORMat[:DATA] command.
·
Returned Value: ASCII values are returned in the form ±1.234567E±123. For
example 13.325 volts would be +1.3325000E+001. Each value is followed by a
comma (,). A line feed (LF) and End-Or-Identify (EOI) follow the last value. The
C-SCPI data type is a string array.
REAL 32, REAL 64 and PACK 64, values are returned in the IEEE-488.2-1987
Definite Length Arbitrary Block Data format. This data return format is explained
in “Arbitrary Block Program Data” on page 180 of this chapter. For REAL 32,
each value is 4 bytes in length (the C-SCPI data type is a float32 array). For
REAL 64 and PACK 64, each value is 8 bytes in length (the C-SCPI data type is a
float64 array).
NOTE Algorithm values which are a positive overvoltage return IEEE +INF and a negative
overvoltage return IEEE -INF (see table on page 230 for actual values for each data
format).
·
Related Commands: DATA:FIFO:COUNT?
·
*RST Condition: FIFO buffer empty
Usage DATA:FIFO:PART? 256 return 256 values from FIFO
[SENSe:]DATA:FIFO:RESet
[SENSe:]DATA:FIFO:RESet clears the FIFO of values. The FIFO counter is reset
to 0.
VT1419A Command Reference
[SENSe]
Chapter 6 265
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com