Specifications

[SENSe:]CHANnel:SETTling
[SENSe:]CHANnel:SETTling <num_samples>,<ch_list> specifies the number of
measurement samples to make on channels in <ch_list>. SENS:CHAN:SETTLING
is used to provide additional settling time only to selected channels that might need
it. See the “Settling Characteristics” discussion on page 101.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
settle_time numeric (int16) 1 to 64 none
ch_list string 100 - 163 none
Comments
·
SENS:CHAN:SETTLING causes each channel specified in <ch_list> that is also
referenced in an algorithm to appear <num_samples> times in the analog input
Scan List. Channels that do not appear in any SENS:CHAN:SETT command will
be entered into the scan list only once when referenced in an algorithm.
·
Since the scan list is limited to 64 entries, an error will be generated if the number
of channels referenced in algorithms plus the additional entries from any
SENS:CHAN:SETTLING commands that coincide with algorithm referenced
channels exceeds 64.
·
The SAMPLE:TIMER command can change the effect of the
SENS:CHAN:SETTLING command since SAMPLE:TIMER changes the amount
of time for each measurement sample.
·
When Accepted: Not while INITiated
·
Related Commands: [SENSe:]CHANnel:SETTling?, SAMPLE:TIMER
·
*RST Condition: SENS:CHAN:SETTLING 1,(@100:163)
Usage SENS:CHAN:SETT 4,(@144,156) settle channels 44 and 56 for 4
measurement periods
VT1419A Command Reference
[SENSe]
Chapter 6 259
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com