Specifications
SAMPle
The SAMPle subsystem provides commands to set and query the interval between
channel measurements (pacing).
Subsystem Syntax SAMPle
:TIMer <interval>
:TIMer?
SAMPle:TIMer
SAMPle:TIMer <interval> sets the time interval between channel measurements.
It is used to provide additional channel settling time. See “Settling Characteristics”
discussion on page 101.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
interval numeric (float32) (string) 1.0E-5 to 16.3825E-3 |
MIN | MAX
seconds
Comments · The minimum <interval> is 10 µs. The resolution for <interval> is 2.5 µs.
·
If the Sample Timer interval multiplied by the number of channels in the specified
Scan List is longer than the Trigger Timer interval, at run time a “Trigger too
fast” error will be generated.
·
The SAMP:TIMER interval can change the effect of the
SENS:CHAN:SETTLING command. ALG:CHAN:SETT specifies the number of
times a channel measurement should be repeated. The total settling time per
channel then is (SAMP:TIMER <interval>)X(<chan_repeats> from
SENS:CHAN:SETT)
·
When Accepted: Not while INITiated
·
Related Commands: SENSE:CHAN:SETTLING, SAMP:TIMER?
·
*RST Condition: Sample Timer for all Channel Lists set to 1.0E-5 seconds.
Usage SAMPLE:TIMER 50E-6 Pace measurements at 50 µs intervals
SAMPle:TIMer?
SAMPle:TIMer? returns the sample timer interval.
Comments
·
Returned Value: Numeric. The C-SCPI type is float32.
256 Chapter 6