Specifications

OUTPut:TYPE?
OUTPut:TYPE? <channel> returns the output drive characteristic for a digital
channel.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel string 132 - 163 none
Comments
·
The <channel> parameter must specify a single channel.
·
If the channel specified is not on a digital SCP, an error will be generated.
·
Returned Value: returns PASS or ACT. The type is string.
·
*RST Condition: returns ACT
OUTPut:VOLTage:AMPLitude
OUTPut:VOLTage:AMPLitude <amplitude>,(@<ch_list>) sets the excitation
voltage on programmable Strain Bridge Completion SCPs pointed to by <ch_list>
(the VT1511A for example). This command is not
used to set output voltage on
SCPs like the VT1531A Voltage Output SCP.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
amplitude numeric (float32) MIN|0|1|2|5|10|MAX none
ch_list channel list (string) 132 - 163 none
Comments
·
To turn off excitation voltage (when using external voltage source) program
<amplitude> to 0.
·
Related Commands: OUTP:VOLT:AMPL?
·
*RST Condition: MIN (0)
Usage OUTP:VOLT:AMPL 5,(@132:135) set excitation voltage for channels 32 - 35
OUTPut:VOLTage:AMPLitude?
OUTPut:VOLTage:AMPLitude? (@<channel>) returns the current setting of
excitation voltage for the channel specified by <channel>. If the channel is not on a
VT1511A SCP, the query will return zero.
Comments
·
The <channel> parameter must specify a single channel.
·
Returned Value: Numeric, one of 0, 1,2,5,or10.C-SCPI type is float32.