Specifications
OUTPut:POLarity
OUTPut:POLarity <select>,(@<ch_list>) sets the polarity on digital output
channels in <ch_list>.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
select discrete (string) NORMal | INVerted none
ch_list string 132 - 163 none
Comments
·
If the channels specified do not support this function, an error will be generated.
·
Related Commands: INPut:POLarity, OUTPut:POLarity?
·
*RST Condition: OUTP:POL NORM for all digital channels
Usage OUTP:POL INV,(@156) invert output logic sense on channel 56
OUTPut:POLarity?
OUTPut:POLarity? (@<channel>) returns the polarity on the digital output
channel in <channel>.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel string 132 - 163 none
Comments
·
The <channel> parameter must specify a single channel.
·
Returned Value: returns one of NORM or INV. The type is string.
OUTPut:SHUNt
OUTPut:SHUNt <enable>,(@<ch_list>) adds shunt resistance to one leg of
bridge on Strain Bridge Completion SCPs. This can be used for diagnostic purposes
and characterization of bridge response.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
enable boolean (uint16) 0|1|ON|OFF none
ch_list channel list (string) 132 - 163 none
Comments
·
If <ch_list> specifies a non strain SCP, a 3007 “Invalid signal conditioning
plug-on” error is generated.
VT1419A Command Reference
OUTPut
248 Chapter 6
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com