Specifications
Usage OUTP:CURR:AMPLITUDE? (@140) Check SCP current set for channel 40
(returns +3.0E-5 or +4.88E-4)
OUTPut:CURRent[:STATe]
OUTPut:CURRent[:STATe] <enable>,(@<ch_list>) enables or disables current
source on channels specified in <ch_list>.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
enable boolean (uint16) 1|0|ON|OFF none
ch_list channel list (string) 132 - 163 none
Comments
·
OUTP:CURR:STAT does not affect a channel’s amplitude setting. A channel that
has been disabled, when re-enabled sources the same current set by the previous
OUTP:CURR:AMPL command.
·
OUTP:CURR:STAT is most commonly used to turn off excitation current to
four-wire resistance (and resistance temperature device) circuits during execution
of CAL:TARE for those channels.
·
When Accepted: Not while INITiated
·
Related Commands: OUTP:CURR:AMPL, CAL:TARE
·
*RST Condition: OUTP:CURR OFF (all channels)
Usage OUTP:CURR OFF,(@140,147) turn off current source channels 40 and 47
OUTPut:CURRent[:STATe]?
OUTPut:CURRent[:STATe]? (@<channel>) returns the state of the Current
Source SCP channel specified by <channel>. If the channel is not on a VT1505A
Current Source SCP, the query will return zero.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel channel list (string) 132 - 163 none
Comments
·
The <channel> parameter must specify a single channel.
·
Returned Value: returns 1 for enabled, 0 for disabled. C-SCPI type is uint16.
·
Related Commands: OUTP:CURR:STATE, OUTP:CURR:AMPL
Usage OUTP:CURR? (@147) query for state of Current SCP channel 47
execute enter statement here enter query value, either 1 or 0
VT1419A Command Reference
OUTPut
Chapter 6 247
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com