Specifications

Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
mem_size numeric to limit of available VME memory none
Comments
·
This command is only available in systems using an Agilent/HP E1405B/06A
command module.
·
The <mem_size> parameter may be specified in decimal, hex (#H), octal (#Q), or
binary(#B).
·
The <mem_size> parameter should be a multiple of four (4) to accommodate 32
bit readings.
·
Related Commands: MEMory subsystem, FORMAT, and FETCH?
·
*RST Condition: MEM:VME:SIZE 0
Usage MEM:VME:SIZE 32768 Allocate 32 kilobytes (kB) of VME memory
to reading storage (8,192 readings)
MEMory:VME:SIZE?
MEMory:VME:SIZE? returns the amount (in bytes) of VME memory allocated to
reading storage.
Comments
·
This command is only available in systems using an Agilent/HP E1405B/06A
command module.
·
Returned Value: Numeric.
·
Related Commands: MEMory subsystem and FETCH?
Usage MEM:VME:SIZE? Returns the number of bytes allocated to
reading storage.
MEMory:VME:STATe
MEMory:VME:STATe <enable> enables or disables use of the VME memory
card as additional reading storage.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
enable boolean (uint16) 1|0|ON|OFF none
Comments
·
This command is only available in systems using an Agilent/HP E1405B/06A
command module.
VT1419A Command Reference
MEMory
Chapter 6 243
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com