Specifications

Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel string 132 - 163 none
Comments
·
The <channel> parameter must specify a single channel.
·
For the VT1536A Isolated Digital I/O SCP, INP:THR:LEV? returns a numeric
value which is one of 5, 12, 24, 48, or 0 (zero) where zero means that the channel
is configured as an output and non-zero values indicate the input threshold in
volts.
Note If an invalid switch combination is set on a VT1536A, INP:THR:LEV? will NOT
return a value and will generate the error 3105 “Invalid SCP switch setting”. This
error will also be generated when *RST is executed. Channels associated with this
error will behave as input channels with unknown threshold levels.
·
Usage: To query the threshold level on the second channel at SCP position 2
send:
INP:THR:LEV? (@117) query 2nd chan on SCP pos. 2
enter statement here returns0|5|12|24|48
VT1419A Command Reference
INPut
240 Chapter 6