Specifications
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
cutoff_freq numeric (float32)
(string)
see comment |
MIN | MAX
Hz
ch_list channel list (string) 132 - 163 none
Comments
·
The <cutoff_freq> parameter may be specified in kilohertz (kHz). A
programmable Filter SCP has a choice of several discrete cutoff frequencies. The
cutoff frequency set will be the one closest to the value specified by
<cutoff_freq>. Refer to Chapter 6 for specific information on the SCP being
programmed.
·
Sending MAX for the <cutoff_freq> selects the SCP’s highest cutoff frequency.
Sending MIN for the <cutoff_freq> selects the SCP’s lowest cutoff frequency. To
disable filtering (the “pass through” mode), execute the INP:FILT:STATE OFF
command.
·
Sending a value greater than the SCP’s highest cutoff frequency or less than the
SCP’s lowest cutoff frequency generates a -222 “Data out of range” error.
·
When Accepted: Not while INITiated
·
Related Commands: INP:FILT:FREQ?, INP:FILT:STAT ON | OFF
·
*RST Condition: set to MIN
Usage INP:FILT:FREQ 100,(@140:143) Set cutoff frequency of 100 Hz for channels
40 through 43
INPUT:FILTER:FREQ 2,(@155) Set cutoff frequency of 2 Hz for channel 55
INPut:FILTer[:LPASs]:FREQuency?
INPut:FILTer[:LPASs]:FREQuency? (@<channel>) returns the cutoff
frequency currently set for <channel>. Non-programmable SCP channels may be
queried to determine their fixed cutoff frequency. If the channel is not on an input
SCP, the query will return zero.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
channel channel list (string) 100 - 163 none
Comments
·
The <channel> parameter must specify a single channel only.
·
This command is for programmable filter SCPs only.
·
Returned Value: Numeric value of Hz as set by the INP:FILT:FREQ command.
The C-SCPI type is float32.
VT1419A Command Reference
INPut
Chapter 6 235
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com