Specifications

Comments
·
Related Commands: DIAG:INT:LINE?
·
Power-on and *RST Condition: DIAG:INT:LINE 1
Usage DIAG:INT:LINE 5 Module will interrupt on VXIbus interrupt line 5
DIAGnostic:INTerrupt[:LINe]?
DIAGnostic:INTerrupt[:LINe]? returns the VXIbus interrupt line that the module
is set to use.
Comments
·
Returned Value: Numeric 0 through 7. The C-SCPI type is int16.
·
Related Commands: DIAG:INT:LINE
Usage DIAG:INT? Enter statement will return 0 through 7
DIAGnostic:OTDetect[:STATe]
DIAGnostic:OTDetect[:STATe] <enable>,(@<ch_list>) enables and disables the
VT1419’s “Open Transducer Detection” capability (OTD). When Open Transducer
Detection is enabled, a very high impedance path connects all SCP channels to a
voltage source greater than 16 volts. If an enabled channel has an open transducer,
the input signal becomes the source voltage and the channel returns an input
over-range value. The value returned is +9.91E+37 (ASCII).
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
enable boolean (uint16) 1|0|ON|OFF none
ch_list channel list (string) 100 - 163 none
Comments
·
Open Transducer Detection is enabled/disabled on a whole Signal Conditioning
Plug-On basis. Selecting any channel on an SCP selects all channels on that SCP
(8 channels per SCP).
·
The DIAG:CAL:TARE:MODE <mode> command affects how OTD is controlled
during the CAL:TARE? operation. When <mode> is set to 0 (the *RST Default),
channels are tare calibrated with their OTD current off. When <mode>is1,
channels that have their OTD current on
(DIAGnostic:OTDetect ON,(@<ch_list>)) are tare calibrated with their OTD
current left on.
·
Related Commands: DIAG:OTDETECT:STATE?, DIAG:CAL:TARE:MODE
·
*RST Condition: DIAG:OTDETECT OFF
VT1419A Command Reference
DIAGnostic
224 Chapter 6