Specifications

set up scan list sequence (ch 0 in this case)
Now run the algorithm that uses the custom reference conversion table
dump reference temp register to FIFO
DIAG:CUST:REF:TEMP
read the diagnostic reference temperature value
SENS:DATA:FIFO?
DIAGnostic:IEEE
DIAGnostic:IEEE <mode> enables (1) or disables (0) IEEE-754 NAN (Not A
Number) and ±INF value outputs. This command was created for the Agilent VEE
platform.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
mode boolean (uint 16) 0 | 1 volts
Comments
·
When <mode> is set to 1, the module can return ±INF and NAN values according
to the IEEE-754 standard. When <mode> is set to 0, the module returns values as
±9.9E37 for INF and 9.91E37 for NAN.
·
Related Commands: DIAG:IEEE?
·
*RST Condition: DIAG:IEEE 1
Usage Set IEEE mode
DIAG:IEEE 1 INF values returned in IEEE standard
DIAGnostic:IEEE?
DIAGnostic:IEEE? returns the currently set IEEE mode.
Comments
·
The C-SCPI type is int16.
·
Related Commands: DIAG:IEEE
·
*RST Condition: DIAG:IEEE 1
DIAGnostic:INTerrupt[:LINe]
DIAGnostic:INTerrupt[:LINe] <intr_line> sets the VXIbus interrupt line the
module will use.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
intr_line numeric (int16) 0 through 7 none
VT1419A Command Reference
DIAGnostic
Chapter 6 223