DIAGnostic:CALibration:TARE[:OTDetect]:MODE
DIAGnostic:CALibration:TARE[:OTDetect]:MODE <mode> sets whether
Open Transducer Detect current will be turned off or left on (the default mode)
during the CAL:TARE operation.
Parameters
Parameter
Name
Parameter
Type
Range of
Values
Default
Units
mode boolean (uint 16) 0 | 1 volts
Comments
·
When <mode> is set to 0 (the *RST Default), channels are tare calibrated with
their OTD current off. When <mode> is 1, channels that have their OTD current
on (DIAGnostic:OTDetect ON,(@<ch_list>)) are tare calibrated with their OTD
current left on.
·
By default (*RST) the CALibration:TARE? command will calibrate all channels
with the OTD circuitry disabled. This is done for two reasons: first, most users do
not leave OTD enabled while taking readings and second, the
CALibration:TARE? operation takes much longer with OTD enabled. However,
for users who intend to take readings with OTD enabled, setting
DIAG:CAL:TARE:OTD:MODE to 1, will force the CAL:TARE? command to
perform calibration with OTD enabled on channels so specified by the user with
the DIAG:OTD command.
·
Related Commands: *CAL?, CAL:SET, DIAG:CAL:SET:MODE?
·
*RST Condition: DIAG:CAL:TARE:MODE 0
Usage configure OTD on during CAL:TARE
DIAG:CAL:TARE:MODE 1 set mode for OTD to stay on
CAL:TARE? start channel tare cal.
DIAGnostic:CALibration:TARE[:OTDetect]:MODE?
DIAGnostic:CALibration:TARE[:OTDetect]:MODE? returns the currently set
mode for controlling Open Transducer Detect current while performing
CAL:TARE? operation.
Comments
·
Returns a 0 when OTD current will be turned off during CAL:TARE?. Returns 1
when OTD current will be left on during CAL:TARE? operation. The C-SCPI
type is int16.
·
Related Commands: DIAG:CAL:TARE:MOD, DIAG:OTD, CAL:TARE?
·
*RST Condition: DIAG:CAL:TARE:MODE 0