Specifications
CALibration:SETup
CALibration:SETup causes the Channel Calibration function to be performed for
every module channel with an analog SCP installed (input or output). The Channel
Calibration function calibrates the A/D Offset and the Gain/Offset for these analog
channels. This calibration is accomplished using internal calibration references. For
more information, see *CAL? on page 311.
Comments
·
CAL:SET performs the same operation as the *CAL? command except that, since
it is not a query command, it doesn’t tie-up the C-SCPI driver waiting for
response data from the instrument. If there are multiple VT1419As in a system,
start a CAL:SET operation on each and then execute a CAL:SET? command to
complete the operation on each instrument.
·
Related Commands: CAL:SETup?, *CAL?
·
When Accepted: Not while INITiated
Usage CAL:SET start SCP Calibration on 1st VT1419A
: start SCP Calibration on more VT1419As
CAL:SET start SCP Calibration on last VT1419A
CAL:SET? query for results from 1st VT1419A
: query for results from more VT1419As
CAL:SET? query for results from last VT1419A
CALibration:SETup?
CALibration:SETup? Returns a value to indicate the success of the last
CAL:SETup or *CAL? operation. CAL:SETup? returns the value only after
the
CAL:SETup operation is complete.
Comments
·
Returned Value:
Value Meaning Further Action
0 Cal OK None
-1 Cal Error Query the Error Queue (SYST:ERR?)
See Error Messages in Appendix B.
-2 No results available No *CAL? or CAL:SETUP done
The C-SCPI type for this returned value is int16.
·
Related Commands: CAL:SETup, *CAL?
Usage see CAL:SETup
VT1419A Command Reference
CALibration
210 Chapter 6