ABORt
The ABORt subsystem is a part of the VT1419A’s trigger system. ABORt resets the
trigger system from its Wait For Trigger state to its Trigger Idle state.
Subsystem Syntax ABORt
CAUTION! ABORT stops execution of a running algorithm. The control output is left at the last
value set by the algorithm. Depending on the process, this uncontrolled situation
could be dangerous. Make certain that the process is in a safe state before halting the
execution of a controlling algorithm.
Comments
·
ABORt does not affect any other settings of the trigger system. When the
INITiate command is sent, the trigger system will respond just as it did before the
ABORt command was sent.
·
Related Commands: INITiate[:IMMediate], TRIGger¼
· *RST Condition: TRIG:SOUR HOLD
Usage ABORT If INITed, goes to Trigger Idle state. If
running algorithms, stops, and goes to
Trigger Idle State.
Chapter 6 185