14 EX7000 Series Introduction
LXI REQUIRED SPECIFICATIONS
TRIGGER RESPONSE TIMES
Minimum Typical Maximum
Trigger Source
LXI WTB
DIO port
LAN event
Alarm
948 µs 1 ms 1.4 ms
948 µs 1 ms 1.4 ms
3.8 ms 4 ms 4.2 ms
700 µs 1 ms 1.8 ms
EVENT RESPONSE TIMES
Minimum Typical Maximum
Event Source → Trigger Destination
LXI WTB → LXI WTB
LXI WTB → DIO port
LXI WTB → LAN event
DIO port → LXI WTB
DIO port → DIO port
DIO port → LAN event
LAN event → LXI WTB
LAN event → DIO port
LAN event → LAN event
Alarm → LXI WTB
Alarm → DIO port
Alarm → LAN event
Operation Complete → LXI WTB
Operation Complete → DIO port
Operation Complete → LAN event
37.5 ns 40 ns 50 ns
54 ns 60 ns 65 ns
1.26 ms 1.4 ms 1.5 ms
56 ns 60 ns 66 ns
26 ns 30 ns 39 ns
1.19 ms 2.0 ms 2.2 ms
3.7 ms 4 ms 4.8 ms
3.7 ms 4 ms 4.8 ms
14.8 ms 16 ms 25 ms
36 ns 36 ns 36 ns
36 ns 36 ns 36 ns
1.2 ms 1.35 ms 1.45 ms
320 µs 500 µs 600 µs
320 µs 500 µs 600 µs
1.8 ms 1.9 ms 2.0 ms
TIMESTAMP ACCURACY
200 µs