Specifications

ANNEX 2—SPECIFICATIONS
X-RAY
X-Ray Unit, General:
Adjustable Patient Table variable from 59 to 89 cm with foot control pedal
located on both front and rear of table to allow easier patient positioning
Longitudinal/lateral electromagnetic and electromechanical locks to provide
safety and security
Floor mounted tube radiography system allow easy positioning with flexibility
and functionality to accurately position the tube in all radiographic procedures
Bucky wall stand (adjustable height) with cassette function using automatic
cassette size sensing
Collimator (manual or automatic) which effectively limits the field of exposure
during radiography and an easily changeable collimator bulb
Frequency generator available in 4 power configurations of 30/ 50/ 65/ 80KW
Self calibrating variable frequency operation delivers higher accuracy, sharter
exposure time and lower patient dose
Control-console includes manual mode technique and parameter control,
anatomical programmer, and automatic exposure control
X-Ray Film Processor, Bench Top:
Rapid processing cycles for fully processed 35x43 cm (14x17 in) radiograph, dry
and ready to read
Automatically feeds exposed films into the processor
Processing tanks molded in polyurethane with three sections for developer, Fixer
And washer
Thermistor/logic circuit, water heat exchanger, and developer recirculation system
control developer temperature to +/- 2°C
Dryer temperature control is thermostatically controlled and easily adjustable
Automatic standby control when no film is processed and reduces water
consumption.
Power requirements: 110volts/220volts, 50/60hz
X-Ray Film Viewer, Double:
White enameled steel casing
Edge to edge viewing
Clear Perspex spring-loaded film clips
Twin Tropical Daylight tubes with illuminated switching per panel
Dimensions: Height x Width = 505 x 724 mm
Power Supply: 110V/220VOLTS, 50/60HZ
ANNEX 2—SPECIFICATIONS
X-RAY
(continued)
Apron Patient Protective:
Flexible lead rubber in tear resistant plastic cover
Sleeveless, half length at back
Fastening with crossed elastic straps with Velcro end