Specifications
• Electrical operated through 3 independent 24V electric motors
• Four antistatic castors with the central locking system
• Arms rests can be adjusted in height and locked
• Power Supply 220V AC, 50/60Hz
Hot Air Sterilizer:
• Automatic hot air sterilizer assures that the temperature can reach in 5/8mins.
• Adjustable capillary thermostat 0- 210°C
• Adjustable manual timer from l0 min. to 120 min.
• Continuous service 24 hrs on 24 hour timer position
• Stainless steel chamber
• 3 stainless steel trays with 1 extractor for trays
• Suitable for laboratory instruments, glasswares, etc.
• Volume capacity: 40 to 50 liters
• Power requirements: 220volts AC, 50/60Hz
ANNEX 2—SPECIFICATIONS
STERILIZATION UNIT
Sterilizer Unit, Steam:
• Cycles: Gravity and Vacuum combination-(121 °C Pre-vacuum, 134°C Pre-
vacuum) Leak: Test and Bowie-Dick Test
• Impact Printer/Recorder of all pertinent cycle data, Data should be printed at the
beginning and end of each cycle and at transition points during the cycle.
• Single Door, Power Door type, Stainless Steel chamber
• Automatic Controller
• Insulated Doors to reduce surface temperature of stainless steel door cover.
• Power supply, 380 3060Hz (to conform on site)
• Built in steam generator
• Chamber Size: 110 liters (or nearest volume available)
• Complete with loading trol1ey, trays and other accessories
Washer, Disinfector:
• Electric Heated
• Floor Mounted
• Microprocessor Controlled
• Chamber Load Capacity 50 liters
• Vertical sliding power door constructed of tinted, tempered glass
• Door opened or closed automatically by using the touch pads on control board if
power failure occurs door can be opened manually
• Pump powered by dual-speed motor
• Automatic detergent injection assures that the right amount of detergent is
dispensed into the water
• Provided special enzymes pretreatment for a highly effective cleaning
• Heating system enables to be achieved high temperature (95 Centigrade) and
maintained throughout the thermal rinse phase
• Drying temperature enables to set 66 Centigrade to 116 Centigrade
• Walls are well insulated for minimum heat loss
• Enable to complete the cleaning process within 26 to 33 minutes
• Complete with loading trolley, trays and other accessories.