Specifications

Provided cylinders (Oxygen, Nitrous Oxide, Air)
Precision Calibrated Flow meters of Oxygen, Medical Compressed Air And
Nitrous Oxide
Flowmeters: Oxygen two tubes- ranges 0.05 to 0.95 L/min and 1.0 to 15.0 L/min,
Nitrous oxide two tubes- ranges 0.05 to 0.95 L/min and 1.0 to 10 L/min and Air
two tubes 0 to 0.95 and to 1 to 15 L/min
System with three Vaporizer
The system can not turn on more than one Vaporizer at a time and Vaporizer and
manifold prevents gas flow through all the vaporizer that are off
Built in Battery in Ventilator, internal rechargeable sealed lead acid
Approximately 30 minutes of operation provided by battery
Complete power failure or sudden gas failure (or low pressure gas) should
indicate visually and audibly
Ventilator to monitor inspiratory, expiration volume and airway pressure
Ventilator screen Height x Width: 7.6 x 15.2 cm
Ventilator operating modes: Volume controlled and pressure limited and pressure
controlled
Ventilator performance ranges: Tidal volume 20 to 1500 ml (Volume control
mode) with incremental setting, Minute volume range 0 to 99.9 L/min, and
pressure inspired range 12 to 60 cm H2O (increments of 1 cm H2O)
Positive End Expiratory Pressure (PEEP) integrated and electronically controlled,
range: off, 4 to 30 cm H2O (increment of 1 cm H2O)
Alarm sittings for Tidal volume, Minute volume, inspired oxygen for Low and
High. Also to be warm setting for Apnea, Low air way pressure, high pressure
Provided Autoclavable sensor
On Castors size diameter: 12.5 cm and brakes: single foot lever locks and unlocks
two front castors
Spare Socket box for other equipment
With Two Shelve and folding side shelf for accessories
Leakage current less than 500 micro ampere
Power Supply: 230Volts +/- 20%, 6A, 60Hz
ANNEX 2—SPECIFICATIONS
OPERATION THEATRE
(continued)
Suction Regulator:
Included regulator, jars, probe (inlet fitting British standard), rail mounted,
and hose.
With overflow safety trap
Internally well designed with the increments in cmH20 & kPa to give accurate
vacuum by precision regulating diaphragm
With two scales -mmHg and -kPa
With three modes Off, On and Max.
With the feature of positive pressure relief valve vents accidental positive
pressure away from patient and system
Two modes high continuous Gauge 0-100KPa
Flow rate 0-80 1/min