Installation guide

6 (of 14) ROXY EC installation guide, edition 1
Unpacking
Advice (for IQ) during unpacking of the individual system parts:
Check the part number, serial number on the s/n label of each unit
and write it down in the IQ document.
Check the contents of accessory kits etc. using the checklist sup-
plied to assure it is complete and undamaged.
Unpack ROXY Potentiostat and follow installation instructions in
user manual and next section of this document.
Never lift the
DECADE II at its front door
, but at its sides.
Unpack the syringe pump and follow installation instructions in
user manual. The pump should placed on the top of the ROXY
Potentiostat.
Electrical connections (Mains, RS232 & trigger ca-
ble)
In case it is necessary to use another mains cable then supplied
make sure that the cable meets all relevant safety and EMC require-
ments (CE, CSA and UL). The manufacturer is not liable if power
cords are used which do not meet the relevant safety and EMC re-
quirements.
Connect the cables provided in the ROXY potentiostat acces-
sory kit: power cable, trigger cable and serial communication
cable to the ROXY Potentiostat as it is shown in the Fig. 3.
Connect the serial cable to the syringe pump as in the Fig. 4.
Connect the equipment to the COM ports in the USB-RS232
converter (Fig. 5):
ROXY Potentiostat – Serial 1
Syringe Pump – Serial 2
Do not connect the converter to the computer yet. It is required to
install the driver first. For instruction how to install the driver of th
USB-to-RS232 converter read further section of the manual.
Connect 1 m long cell cable (250.0125) inside the potentiostat.