Installation guide

4 (of 14) ROXY EC installation guide, edition 1
Requirements
The size requirement for an ROXY EC system are as follows:
Description Dimensions Weight
ROXY Potentiostat 44 (D) x 22 (W) x 44 (H) cm 14 kg
Syringe pump 11.4 (D) x 23 (W) x 14 (H) cm 2.2 kg
Prior to installation, it is advised to make enough space available
on the lab bench/table.
Ensure that the system is installed on a stabile, flat and smooth
surface and that the table has enough mechanical strength to
hold the weight of the system. A full-dressed system can weigh
up to 17 kg.
Check that the environmental requirements for the ROXY EC
system are met in the laboratory:
Align the system components as indicated in the Fig. 1. Make
sure that the air flow of fans in the equipment is not obstructed
(ROXY Potentiostat: bottom) because this could impair the cool-
ing capability.
Parameter Requirement
Operating temperature 10 – 35
o
C
Operating humidity 20 – 80%, non-condensing