8 VR NxT| Instruction Manual 1.0 02-2019|VRNXTinstr1EN|© 2019 Menarini Silicon Biosystems SpA
Safety
Risks for the samples
l Damaged tubes must not be used.
l Use only liquids and sample containers indicated in this manual. Buffers and tubes different
from those indicated in this manual are not suitable and should not be used.
l Handle samples and closed holder with samples inside with care to avoid creating bubbles
and nullifying the centrifuge effect.
l Do not process samples with bubbles. In case of bubbles, the output sample volume is not
guaranteed.
l For the supernatant to be trapped correctly in the cap, the input sample volume must not
exceed 200 µl.
l Do not use the instrument and its components if not intact.
l Do not apply labels on the tube containing the sample.
l Check the presence of the holder clips. If they are missing, do not use the holder.
l To guarantee correct sample processing, keep the closed holder perpendicular to the plane
(see "Position the closed holder" on page21).
l Do not process samples having a density higher than 1.2 g/mL.
l Use the swing bucket centrifuge only when prescribed. Do not use fixed angle centrifuge.
l Do not use accessories whose maximum service life has been exceeded.
l If unusual noise comes from the instrument, the instrument must be immediately stopped (see
"Stop the volume reduction process" on page31).
Product liability
The product is the result of the professional experience gained by Menarini Silicon Biosystems.
It must be installed and used with the indicated and authorized accessories in compliance with
this instruction manual.
Liability for any property damage or personal injury is transferred to the operator in the following
cases:
l use of the instrument for purposes other than that indicated in "Intended use" on the facing
page
l use of accessories and consumables other than those authorized by Menarini Silicon
Biosystems
l attempts to open and repair the instrument autonomously or with the help of personnel not
authorized by Menarini Silicon Biosystems
l instrument alterations
Intended users
VR NxT is intended to be used by laboratory technicians, biologists, biotechnologists and
researchers with a deep knowledge of the typical laboratory processes.