Troubleshooting and errors
VR NxT| Instruction Manual 1.0 02-2019|VRNXTinstr1EN|© 2019 Menarini Silicon Biosystems SpA 39
Code Cause Solution
ED028 Rotor not aligned 1. Ensure no foreign body is in the rotor chamber.
2. Turn off and on the instrument.
3. If the problem persists, contact Technical Support.
ED029 Closed holder not correctly inserted 1. Remove and check the closed holder.
2. Insert the closed holder again in its housing and push it down until
it clicks.
ED030 Lid lock/unlock failure 1. Close the rotor chamber.
2. Turn off and on the instrument.
3. If the problem persists, contact the technical support.
ED031 At the end of the process, the closed
holder housing has not reached the
right position.
1. Turn off and on the instrument.
2. If the problem persists, contact Technical Support.
ED032 RFID comunication problem 1. Turn off and on the instrument.
2. If the problem persists, contact Technical Support.
ED033 The samples have not been correctly
processed due to an out of range rotor
speed during the process.
Restart sample processing or perform the recovery procedure (see
"Recover the samples" on page32).