Maintenance of the instrument
VR NxT| Instruction Manual 1.0 02-2019|VRNXTinstr1EN|© 2019 Menarini Silicon Biosystems SpA 35
5. Open the rotor chamber and clean the four rotor holes with the hole brush. Reagents are not required.
6. Close the rotor chamber.
7. To move the rotor back in its correct position, connect the instrument to the power supply and switch it on.
Clean the internal parts of the rotor chamber
1. Switch off the instrument and disconnect it from the power supply.
2. Wear gloves.
3. Open the lid.
4. Remove any foreign body from the rotor chamber.
5. Put the DNA AWAY on a swab large head.
6. Use the swab large head to clean the bottom of the closed holder housing and the internal surfaces of the rotor
chamber.
7. Prepare a solution made of water and a 70% Ethanol solution and put it on another swab large head.
8. Use this swab large head to remove any residue of DNA AWAY from the bottom of the closed holder housing and
the internal surfaces of the rotor chamber.
9. Let the instrument dry.
Clean the holder
NOTICE: The holder is not waterproof. Do not submerge in liquids.
1. Wear gloves.
2. Put the DNA AWAY on a swab small head.