® FastStart IF Patient Operating Manual
VQ Wholesale is not liable for any misuse or misunderstanding of the FastStart® IF product or patient manual. Please call your local representative if any additional assistance is required regarding this product and its patient instructions. Irvine, CA 92614 This patient manual is proprietary and confidential.
Table of Contents FastStart® IF Page System Components . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2 Letter to the Patient . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 Precautions and Prescription Information. . . . . . . . . . . . . . . . . . . . . . . . . . 4 Controls and Features . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6 Proper Skin Care . . . . . . . . . . . . . . . . . . . . . .
System Components The FastStart® IF stimulation system is operated with: · · · · One (1) FastStart IF stimulator One (1) package of electrodes One (1) lead wire (4 lead) Four (4) AA alkaline batteries, or one (1) battery pack In addition your system comes with: · · · One (1) carrying case One (1) operating manual One (1) AC Adapter Display Screen Lead Wire Receptacle A/C Line Cord Receptacle Key Pad Belt Clip A/C Line Cord Battery Compartment Lead Wires FRONT ELECTRODES 2 BACK BATTERY PACK
Letter to the Patient Your physician has selected the FastStart® IF device because it is the most applicable to your specific need. Your FastStart IF is a portable stimulator that generates small pulses of electrical current that are delivered through lead cables and the electrodes you place on your skin. These electrical pulses pass through your skin and activate underlying nerves. At lower levels of stimulation, the FastStart IF stimulates sensory nerves, producing a tingling sensation.
Precautions and Prescription Information NOTE: Please read the following prescription information carefully before using your FastStart® IF. If you have any questions on this information, consult with your physician of therapist before proceeding. CAUTION: Federal (U.S.A.) law restricts this device to sale by or on the order of a practitioner licensed by law of the State in which he/she practices to use or order the use of the device.
Precautions · Safety for use during pregnancy has not been established. · Caution should be used for patients with suspected or diagnosed heart problems. · Caution should be used for patients with suspected or diagnosed epilepsy.
Controls and Features The FastStart® IF can be operated on a tabletop by folding out the belt clip and setting on top of the table. You can also attach the unit to your belt by folding the clip against the unit and fastening the clip over the top of a belt or pants. ON/OFF Button Press the On/Off button when you are ready to begin your therapy session. The FastStart IF always turns on with 0 intensity.
· Lower right is your output voltage. · In between the rows there are flashing bars to indicate the current is flowing. · Low battery indicator to the left of the preset display. · Pause indicator to the right of the remaining time display. Low Battery Indicator This will flash on when the batteries are nearly depleted, indicating you have approximately 5-30 minutes of battery life left, depending on your parameter settings. Try to replace them as soon as possible.
Proper Skin Care CAUTION: DO NOT apply lotion, oils or any other ointments to skin prior to the application of electrodes, unless directed to by your clinician or your dealer representative. Topical agents may increase your risk of skin irritation. DO NOT clean skin with anything other than soap and water. DO NOT clean skin with alcohol, alcohol-based cleaners or Iodine-based antiseptic. This may interfere with proper conduction and increase the risk of skin irritation.
Treatment Sessions 1. Insert (4) AA batteries or (1) battery pack into the FastStart® IF as shown in the battery compartment and then replace the battery door. If desired, use the AC adapter that came with your FastStart IF instead of batteries. 2. Prepare the skin: Wash the electrode application sites with mild soap and water then dry skin thoroughly. Remove any excess body hair with a pair of scissors, being careful not to cut your skin. 3. Connect the lead cables to the electrodes.
Care and Maintenance CAUTION: DO NOT immerse the FastStart® IF stimulator, AC adapter, or batteries in water. Cleaning Maintenance, care and cleaning of your FastStart IF are quick and easy procedures when these instructions are followed. Clean the outside of the FastStart IF casing as needed with a damp cloth and mild soap. Never immerse your stimulator or batteries in water, alcohol or other fluids since this could seriously damage the internal electronics.
Technical Data and Specifications Physical Power Source: Size: Weight: One (1) FastStart battery pack or FastStart AC Adapter 5.25" x 2.8" x 1.75" (including belt clip) 10.8 oz. with batteries Electrode Use with any electrode approved for use with electrical stimulators Electrical Tolerances: All electrical specifications are ± 10% into a 500 Ohm load impedance for each channel and using the AC Adapter, unless stated otherwise.
Technical Data and Specifications (cont.
Waveform Output of Carrier Frequency Figure 1 30 25 20 15 amplitude (volts) 10 5 0 -5 50 100 150 200 time (microseconds) 300 350 400 450 -10 -15 -20 -25 -30 load = 500 Ohms load = 2000 Ohms load = 10000 Ohms Figure 1: Electric output of carrier frequency with various loads with Amplitude Control set at 50% of maximum.
IF Mode Freq. 160% of User Setting Mode: 1|1 User Setting 60% of User Setting time (each hash mark = 1 second) Freq. 160% of User Setting Mode: 6|6 User Setting 60% of User Setting time (each hash mark = 1 second) Freq. 160% of User Setting Mode: 6/6 User Setting 60% of User Setting time (each hash mark = 1 second) Figure 2: 14 Electric output of carrier frequency on 1000 ohm resistive load with Amplitude Control set at 50% of maximum. Note constancy of voltage relative to Figure 1.
NMES Mode Freq. User Setting Mode: C Off time (each hash mark = 1 second) Freq. 160% of User Setting Mode: 1|1 User Setting 60% of User Setting Off time (each hash mark = 1 second) 160% of User Setting Freq. Mode: 6|6 Off time (each hash mark = 1 second) Freq. 160% of User Setting Mode: 6/6 User Setting 60% of User Setting Off time (each hash mark = 1 second) Figure 3: Voltage (top) and current (bottom) outputs of carrier frequency on 1K AAMI load (mimicking skin impedance).
16 Figure 4: Electrical output during bipolar (2-electrode) stimulation measured across the capacitive element of the AAMI load. This shows the interaction of the carrier frequency and adjustable frequency which takes place internal to the stimulator. The interference frequency is evident as an “envelope” of stimulation occurring at 100 Hz. This figure is provided to illustrate interferential stimulation, but does not correspond exactly to the output which an oscilloscope would record with a skin load.
Warranty VQ Wholesale warrants the FastStart® IF to be free from electrical circuit defects in workmanship and materials for a period of three (3) years upon shipment. The casing and belt clip are warranted to be free from defects in workmanship and materials for one (1) year upon shipment. VQ Wholesale will repair or replace, at its facilities, any units found to have become defective within the warranty period.