Operation Manual
16
ENGLISH
NOTE: The electrodes supplied with this Avid IF 2 have been designed
and authorized specifically for use with this product and must not be
manipulated, cut, or altered in any way as this may affect the even
distribution of stimulation and/or its intensity beyond the recommended
0.25W per cm
2
maximum.
NOTE: Do not place electrodes on cut, broken or irritated skin. If irri-
tation occurs, consult with your physician or therapist before changing
your placement sites.
To apply self-adhesive, reusable electrodes:
Attach the electrodes to the lead wires provided with the Avid IF 2.
Grasp an electrode by the edge and gently peel away the electrode
from its plastic liner.
Apply the electrodes to the affected area by firmly pressing the elec-
trode against the skin, checking all edges for proper adhesion. Your
physician or therapist will show you the application sites for treatment.
Always make sure electrodes are placed at least one inch apart from
each other.
To remove the electrodes:
Grasp the electrode by the edge and gently remove it from the skin.
Return it to the plastic liner.
Replace the electrodes on the plastic liner in the resealable
pouch provided.
CAUTION: Never remove the electrode by pulling on the wire, as
this may damage your electrode.
CAUTION: Be careful when using the Avid IF 2 during sleep, as
wires may tangle and become loose.
Electrode Use and Care
WARNING: This product should be used only under the direct super-vision of
a registered healthcare practitioner. Product use should be in accordance with
the directions and specifications of the stimulation device. Improper use of this
or any electrode may result in a patient burn. Due to the potential for causing
skin burns, devices must be limited to 0.25W per cm
2
. Electrodes should not be
placed on the chest in such a configuration that would allow electrical current
delivered by the Avid IF 2 to flow through the chest as it may cause cardiac
arrhythmia.