Specifications

18
MAINTENANCE
CALIBRATION
Calibration Function
1. The initial WC (initial weight on load cell) reading is an arbitrary value.
2. Press SET TARE and the WC displays 0 signifying the TARE has been selected.
3. Install calibration weight in weigh chamber.
4. WC will display the uncorrected weight.
5. Press blue “Cal lbs” and enter value of calibration weight.
6. Press CALIBRATE.
7. WC now displays the corrected weight.
8. % ADJUST shows the amount of correction applied (positive or negative).
NOTE: Each time the screen is opened, WC will contain an arbitrary value until TARE is pressed. Setting
TARE will then initiate the calibration process.
SETTING INSTRUCTIONS FOR LEVEL SENSORS
LEVEL SENSOR ADJUSTMENT
Level Sensor Adjustment Procedure
L-R Systems A125-L01 (KN0006 2-wire AC Normally Closed) High or Low level sensor
L-R Systems A125-L02 (KN0005 2-wire AC Normally Open) Low-Low level sensor
Note: There is only one N/O sensor in your blender eg: Mixer low level
With target product absent, press the push button until the green LED flashes (5 second Maximum) and then
release the button. After a short period, the green LED becomes constant.
The unit is now ready for operation.