User`s guide
Chapter 1 Introducing the Voyager™ Biospectrometry™ Workstations
1-26 PerSeptive Biosystems
1
Vacuum pumps
Two vacuum pumps create the vacuum environment:
• Fore pump—Creates a vacuum in the sample loading 
chamber, creates a lower-than-atmospheric-pressure 
condition before the turbo pump starts, and provides 
backing pressure to the turbo pump.
• Turbo pump—Creates a high vacuum condition in the 
main source chamber.
Vacuum is maintained in the chambers by valves that isolate 
the chambers.
Vacuum gauges
The Voyager-DE Biospectrometry Workstation includes two 
vacuum gauges:
• BA1 (Bayard-Alpert Gauge)—Monitors pressure in the 
main source chamber. Called BA1 on Vacuum Gauge 
Panel. 
• TC2—Monitors pressure in the sample chamber.
Readings from the vacuum gauges are displayed:
• On the Vacuum Gauge Panel. See Section 1.5.4, 
Vacuum Gauge Panel. 
• In the System Status Control page in the Instrument 
Control Panel. See Section 2.9, Checking System Status.