Obtaining Good Spectra in Continuous Extraction Mode
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide H-15
H
Fine-tuning
threshold
When you determine the high and low setting for the sample 
class/matrix you are analyzing, you can fine-tune the threshold 
by setting the laser intensity midway between the high 
setting and low setting determined above.
If signal is present when you decrease the laser, assume 
that this is the new high setting. If signal is not present 
when you decrease the laser, assume that this is the new 
low setting. 
Continue adjusting until the difference between high and 
low settings is less than 10 laser counts.
NOTE: If the difference between the high and low setting is 
equal to the laser step size, decrease the laser step size in 
the Hardware Configuration by a factor of 2. 
Figure H-6 shows a spectrum generated with an acceptable 
laser intensity.
Figure H-6 Spectrum at Lower Laser Setting