User`s guide
Acquiring PSD Data with Standard Instrument Settings (.BIC) Files
Voyager
™
 Biospectrometry
™
 Workstation User’s Guide 7-27
7
NOTE: The calibration you specify on the PSD Acquisition 
Settings control page is used to determine the values for 
α 
, 
β 
, and 
γ
 needed for the PSD calibration equation 
(described on page 7-8). The calibration you specified on 
the Instrument Settings control page in step 4 is used to 
determine the value for 
t
p 
(precursor ion flight time) 
needed for the PSD calibration equation. 
NOTE: If default PSD calibration yields acceptable 
mass accuracy for your application, an external 
calibration is not required.
NOTE: PSD .CAL files are named with the same 
extension as reflector or linear .CAL files. If you select a 
reflector or linear .CAL file when in PSD mode, an error 
message is generated.